Instructions(How to maintain the contra angle of C-SAILOR)

As we all know, the cleaning and maintenance of contra angle of implant machine are very important. Correct cleaning and maintenance can effectively prolong the life of contra angle, improve the efficiency of treatment, and increase the trust of patients. More importantly, it can effectively prevent cross-infection. Today, we will teach you how to clean and maintain the contra angle of C-SAILOR.

instruction926 Instructions How to
maintain the contra angle of C-SAILOR
instruction926 Pre-Cleaning 1.Remove the bur, scrub the surface of the handpiece with a piece of cloth with 60%-90% concentration of ethanol or disinfectant.
instruction926 2.Dip the head into the water in depth of about 20mm, and activate the contra angle clockwise at speed of 300RPM for more than 15 seconds. 3.Then make the head ratate anticlockwise at the speed of 300RPM for more than 15 seconds
instruction926 Note:
1.At the suturing stage of implanting surgery, please let the nurse remove the implanting handpiece and do the pre-cleaning in case that the blood outside and inside of the handpiece clot.
2. The water source requirement: 1) Water temperature: 20℃-38℃ ; 2) Drinkable water.
Manual cleaning
instruction926 1.Pull out the contra angle from the handle of the
motor in order to clean.
instruction926 2.Brush and clean the surface of the handpiece with a brush in the flowing water for 2-3 minutes. 3.Clean the inner hole of the cartridge with ø0.7mm interdental brush in the flowing water. Note:
1.The water source requirement: 1) Water temperature: 20℃-38℃ ; 2) Drinkable water; 3) Water yield: more than 3.5 litters per minute.
instruction926 Oil
Cleaning
& lubrication
(1)Oil cleaning and lubrication of detachable contra angle 1.Grip the tail case and twirl the head anticlockwise then detach it
instruction926 2.Demount the shaft and handle
respectively and put them in clean
container.
3.Put the nozzle tip into the inner hole of the head (with needle) and inject the lubricant for at least 3 seconds until the lubricant discharged is clean.
instruction926 4. Put the nozzle tip into the inner
hole of the head (with shaft) and
inject the lubricant for at least 3s
until the lubricant discharged is clean.
5.Aim the nozzle tip at the bearing of the shaft and inject the lubricant for at least 3 seconds until the lubricant discharged is clean.
instruction926 6. Put the nozzle tip into the inner hole
of the contra part and inject the
lubricant for at least 3 seconds until the
lubricant discharged is clean.
7. Put the nozzle tip into the inner hole of the handle and inject the lubricant for at least 3 seconds until the lubricant discharged is clean.
instruction926 2)Oil cleaning and lubrication of undetachable contra angle. 1. Put the nozzle tip into the inner hole of the head (with needle) and inject the lubricant for at least 3 seconds until the lubricant discharged is clean. 2. Put the nozzle tip into the inner hole
of the handle and inject the lubricant for
at least 3 seconds until the lubricant
discharged is clean.
instruction926 High Temperature
Sterilization
1. Put the handpiece into the special
sterilization bag, seal and put the bag
on the sterilization tray.
instruction926 2. Requirement of high temperature and high pressure sterilization: 134℃, no less than 5 minutes.
instruction926 Examination
Before Use
1.Swirl the head cap to confirm whether
it is loose or not.
instruction926 2.Confirm whether the bur and withdrawer are abnormal, whether the needle is firmly fixed. 3. Confirm whether the water yield is
abundant.
instruction926 4.Activate the handpiece at the highest speed allowed for 1 minute. 5.Shut down the handpiece and examine
whether the head is overheated.